36 CFR Part 1195 Standards for Accessible Medical Diagnostic Equipment - Preamble
M302.3.2 Leg Supports (Section-by-Section Analysis)
The MDE NPRM did not propose to require stirrups to provide a method of supporting, positioning, and securing the patient’s legs for diagnostic equipment used by patients in the seated position. However, in response to question 23, on whether diagnostic equipment used by patients in a seated position that provide stirrups should have to provide such support, the Board received six comments. NPRM, 77 FR at 6926. All six commenters concurred that when stirrups are provided for use with diagnostic equipment used by patients in the seated position, a method must be provided for supporting, positioning, and securing the patient’s legs. The MDE Advisory Committee did not address this provision.
The Access Board concurs with the commenters, and the final rule requires that where stirrups are provided on seated diagnostic equipment, leg supports must also be provided and must comply with the technical requirements for leg supports in M305.4. This will ensure that patients with limited leg strength and control will be able to keep their legs in the appropriate position for examination.