M303.2 Wheelchair Spaces.
Wheelchair spaces complying with M303.2 shall be provided at diagnostic equipment.
M303.2.1 Orientation.
Wheelchair spaces shall be designed so that a patient seated in a wheelchair orients in the same direction that a patient not seated in a wheelchair orients when the diagnostic equipment is in use.
M303.2.2 Width.
Wheelchair spaces shall be 36 inches (915 mm) wide minimum.
EXCEPTION: Wheelchair spaces located on raised platforms shall be permitted to be 32 inches (815 mm) wide minimum to a height of 4 inches (100 mm) measured from the platform surface.
M303.2.3 Depth.
The depth of wheelchair spaces shall comply with M303.2.3.
M303.2.3.1 Front or Rear Entry.
Where wheelchair space entry and exit is provided at only one end (front or rear) the wheelchair space shall be 48 inches (1220 mm) deep minimum.
M303.2.3.2 Pass Through Entry.
Where wheelchair space entry and exit permits pass through from one end to the other, the wheelchair space shall be 40 inches deep (1015 mm) minimum.
M303.2.3.3 Side Entry.
Where wheelchair space entry is only from the side, the wheelchair space shall be 60 inches (1525 mm) deep minimum.
M303.2.4 Equipment Clearances.
Where wheelchair spaces are entered from the rear and includes space beneath components, wheelchair spaces shall include knee and toe clearances complying with M303.2.4.1 for breast platforms and M303.2.4.2 for all other equipment.
M303.2.4.1 Breast Platforms.
Wheelchair spaces beneath breast platforms shall comply with M303.2.4.1.
M303.2.4.1.1 Depth.
Wheelchair spaces shall include knee and toe clearance 25 inches (635 mm) deep minimum and 28 inches (710 mm) deep maximum.
M303.2.4.1.2 Height.
Wheelchair spaces shall include toe clearance 9 inches (230 mm) high minimum above the floor measured to a depth of 6 inches (150 mm) maximum from the toe end of the wheelchair space. Knee clearance shall be provided at a depth of 19 inches (485 mm) minimum and 22 inches (560 mm) maximum at 9 inches (230 mm) above the floor and at a depth of 16 inches (405 mm) minimum at 27 inches (685 mm) above the floor measured from the leading edge of the breast platform. Between 9 inches (230 mm) and 27 inches (685 mm) above the floor, the knee clearance shall be permitted to reduce at a rate of 1 inch (25 mm) in depth for every 6 inches (150 mm) in height.
EXCEPTION: Components shall be permitted to extend into the wheelchair space at a height of 1 ½ inches (38 mm) maximum between 17 inches (430 mm) minimum and 25 inches (635 mm) maximum in depth measured from the leading edge of the breast platform. From 25 inches (635 mm) to 28 inches (710 mm) in depth the height of a component above 1 ½ inches (38 mm) shall be beveled at a rate of 2.5:3 maximum.
M303.2.4.2 Other Equipment.
Wheelchair spaces beneath diagnostic equipment other than breast platforms shall comply with M303.2.4.2.
M303.2.4.2.1 Depth.
Wheelchair spaces shall include knee and toe clearance 17 inches (430 mm) deep minimum and 25 inches (635 mm) deep maximum.
M303.2.4.2.2 Height.
Wheelchair spaces shall include toe clearance 9 inches (230 mm) high minimum above the floor measured to a depth of 6 inches (150 mm) maximum measured from the toe end of the wheelchair space. Knee clearance shall be provided at a depth of 11 inches (280 mm) minimum and 25 inches (635 mm) maximum at 9 inches (230 mm) above the floor and at a depth of 8 inches (205 mm) minimum at 27 inches (685 mm) above the floor measured from the leading edge of the equipment. Between 9 inches (230 mm) and 27 inches (685 mm) above the floor, the knee clearance shall be permitted to reduce at a rate of 1 inch (25 mm) in depth for every 6 inches (150 mm) in height.
M303.2.5 Surfaces.
Wheelchair space surfaces shall not slope more than 1:48 in any direction.
M303.2.6 Edge Protection.
Where wheelchair spaces are provided on a platform raised more than 1 ½ inches (38 mm) in height, edge protection 2 inches (51 mm) high minimum measured from the surface of the platform shall be provided on each side not providing entry to or exit from the equipment.
User Comments/Questions
Add Comment/Question