36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM
M303.3.3 Ramped.
Changes in level greater than ½ inch (13 mm) high shall be ramped and shall comply with M303.3.3.
M303.3.3.1 Running Slope.
Ramp runs shall have a running slope not steeper than 1:12.
M303.3.3.2 Cross Slope.
The cross slope of ramp runs shall not be steeper than 1:48.
M303.3.3.3 Clear Width.
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
M303.3.3.4 Edge Protection.
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side.
M303.3.3.5 Handrails.
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on each side.