36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM

This is an NPRM for the Proposed MDE Standards (2012). Click here to view the Final MDE Standards (2017).

M305.2 Transfer Supports.

Transfer supports shall comply with M305.2.

M305.2.1 Location.

Transfer supports shall be located within reach of the transfer surface and shall not obstruct transfer onto or off of the surface when in position.

M305.2.2 Structural Strength.

Transfer supports and their connections shall be capable of resisting vertical and horizontal forces of 250 pounds (1,112 N) applied at all points on the transfer support.

M305.2.3 Fittings.

Transfer supports shall not rotate within their fittings.