36 CFR Part 1195 Standards for Accessible Medical Diagnostic Equipment

M305.2.2.1 End Transfer Supports.

Transfer supports for transfer surfaces complying with M301.2.3.1 and M305.2.2.1 shall be 15 inches (380 mm) long minimum. Transfer supports shall be positioned along 13 ½ inches (345 mm) minimum of the depth of the transfer surface.