M305 Supports
M305.1 General.
Supports shall comply with M305.
M305.2 Transfer Supports.
Transfer supports shall comply with M305.2.
M305.2.1 Location.
Transfer supports shall comply with M305.2.1.1 or M305.2.1.2 and shall be located 1 ½ inches (38 mm) maximum measured horizontally from the plane defined by the nearest edge of the transfer surface.
EXCEPTION: Where the support folds, collapses, or articulates, the transfer support shall be permitted to be located 3 inches (75 mm) maximum from the plane defined by the nearest edge of the transfer surface.
M305.2.1.1 End Transfer Supports.
Transfer supports for transfer surfaces complying with M301.2.3.1 and M302.2 shall be located on the short side (length) opposite the transfer side.
M305.2.1.2 Side Transfer Supports.
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be capable of supporting transfer on each side of the transfer surface.
M305.2.2 Length.
The length of transfer supports shall comply with M305.2.2.1 or M305.2.2.2.
M305.2.2.1 End Transfer Supports.
Transfer supports for transfer surfaces complying with M301.2.3.1 and M305.2.2.1 shall be 15 inches (380 mm) long minimum. Transfer supports shall be positioned along 13 ½ inches (345 mm) minimum of the depth of the transfer surface.
M305.2.2.2 Side Transfer Supports.
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be 28 inches (710 mm) long minimum and shall be positioned along the width of transfer surfaces.
EXCEPTIONS: 1. Where transfer surfaces are part of an articulating surface, the support shall be permitted to be 15 inches (380 mm) long minimum.
2. Where the width of an imaging bed is more than 24 inches (533 mm), transfer supports shall be permitted to be 12 inches (305 mm) long minimum.
M305.2.3 Height.
During use, the tops of transfer support gripping surfaces shall be 6 inches (150 mm) minimum and 19 inches (485 mm) maximum higher than the top of the associated uncompressed transfer surface.
EXCEPTION: Where the width of the transfer surface for imaging beds exceed 24 inches (610 mm), the tops of the gripping surfaces shall be permitted to be 3 inches (75 mm) minimum and 6 inches (150 mm) maximum higher than the top of the associated uncompressed transfer surface.
M305.2.4 Cross Section.
Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2.
M305.2.4.1 Circular Cross Section.
Transfer supports with circular cross sections shall have an outside diameter of 1 ¼ inches (32 mm) minimum and 2 inches (51 mm) maximum.
M305.2.4.2 Non-Circular Cross Section.
Transfer supports with non-circular cross sections shall have a cross-section dimension of 2 inches (51 mm) maximum and a perimeter dimension of 4 inches (100 mm) minimum and 4.8 inches (120 mm) maximum.
M305.2.5 Surface Hazards.
Transfer supports and surfaces adjacent to transfer supports shall be free of sharp or abrasive components and shall have eased edges.
M305.2.6 Gripping Surface.
Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides. The bottoms of transfer support gripping surfaces shall not be obstructed for more than 20 percent of their length.
M305.2.7 Clearance.
Clearance between the transfer support gripping surface and adjacent surfaces or obstructions shall be 1 ½ inches (38 mm) minimum.
M305.2.8 Fittings.
Transfer supports shall not rotate within their fittings when in place for transfer.
M305.3 Standing Supports.
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3.
M305.3.1 Length.
The length of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve. Horizontal standing support gripping surfaces shall comply with M305.3.1.1, horizontal standing support gripping surfaces on diagnostic equipment containing a wheelchair space shall comply with M305.3.1.2 and, vertical standing support gripping surfaces shall comply with M305.3.1.3.
M305.3.1.1 Horizontal Position.
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply with M305.3.1.2.
M305.3.1.2 Diagnostic Equipment Containing a Wheelchair Space.
On diagnostic equipment containing wheelchair spaces with one entry that also serves as the exit, the length of the gripping surface of horizontal standing supports shall be equal to or greater than 80 percent of the overall length of the platform. On diagnostic equipment containing a wheelchair space and permitting pass-through from one end to the other, the length of the gripping surface on horizontal standing supports shall be at least equal to the length of the platform.
M305.3.1.3 Vertical Position.
The length of the gripping surface on vertical standing supports shall be 18 inches (455 mm) minimum.
M305.3.2 Height.
The height of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve. Horizontal standing support gripping surfaces shall comply with M305.3.2.1 and vertical standing support gripping surfaces shall comply with M305.3.2.2.
M305.3.2.1 Horizontal Position.
The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface.
M305.3.2.2 Vertical Position.
The height of the lowest end of the gripping surface on vertical standing supports shall be 34 inches (865 mm) minimum and 37 inches (940 mm) maximum above the standing surface.
M305.3.3 Fittings.
Standing supports shall not rotate within their fittings.
M305.4 Leg Supports.
Leg supports shall provide a method of supporting, positioning, and securing the patient’s legs.
M305.5 Head and Back Support.
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided. Where the incline of the back support can be modified while in use, head and back support shall be provided throughout the entire range of the incline.