36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment - Example Applications of Proposed Standards
Figure M2: M301.4 Lift Compatibility
(at Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position)
1. 6 inches (150 mm) high min. clear above finished floor
2. patient lift
3. examination surface
4. clearance must extend 36 inches (915 mm) min. deep from the edge of the examination surface
5. base is permitted to increase in width at a rate of 1 inch (25 mm) in width for each 3 inches (75 mm) in depth within clearance
M301.4.1 Clearance in Base
the clearances required for lift compatibility permits equipment to be usable with a patient portable floor lift
M301.4.2 Clearance Around Base
1. examination surface
2. patient lift
3. 6 inches (150 mm) high min. clear above finished floor within clearance
4. 8 inches (205 mm) max.
User Comments/Questions