36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment - Example Applications of Proposed Standards

Figure M3: Overview of M302 Diagnostic Equipment for Use by a Patient in a Seated Position

Three-dimensional view of a height adjustable exam chair showing features and minimum dimensions required by the proposed Standards.

1. when adjustable, head and back support provided throughout the entire range of the incline (M302.3.3)

2. transfer surface 21 inches (760 mm) wide min. and 15 inches (381 mm) deep min. (M302.2.2)

3. armrest folds up to permit unobstructed transfer (M302.2.3 EXCEPTION)

4. transfer surface 17 inches (430 mm) min. and 19 inches (485 mm) max. above floor (M302.2.1), when not needed to facilitate transfer, the transfer surface may be positioned above or below the height range (Advisory M302.2.1)

5. required armrest serves as transfer support within reach of transfer surface (M302.3.1, M302.3.2, and M305.2.1)

6. transfer support resists vertical and horizontal forces of 250 lbs at all points and does not rotate within it's fittings (M305.2.2 and M305.2.3)

7. one short side (depth) and one long side (width) of the transfer surface permit unobstructed transfer from a mobility device (M302.2.3)

8. base permits clearance around base for a patient portable floor lift (M302.4 and M302.4.2)

Woman transferring from a manual wheelchair positioned at an angle to the side of the exam chair shown in the overview image.

transfer to short side (depth) of transfer surface

Man transferring from a manual wheelchair positioned perpendicular to the front of the exam chair shown in the overview image.

transfer to long side (width) of transfer surface